Journal of Applied Physiology, Vol 38, Issue 3 531-537, Copyright © 1975 by American Physiological Society
ARTICLES |
K. E. Finucane and J. Mead
A method for obtaining a continuous estimate of alveolar pressure (PAlv) during periodic flow is described; it was developed to improve the precision of measurements of airway and respiratory tissue impedance using the improved resolution of relatively high-frequency (approximately 5 Hz) singlas. The respiratory system was modulated with a piston pump, and lung volume and the volume change due to compression and expansion of alveolar gas were measured plethysmorgraphically; these signals and an analog divider were used to obtain a continuous solution of Boyle's law during flow. The plethysmorgraph was of the "flow" type; with it volume changes at frequencies up to 10 Hz and with rates of change up to 6 l/s were measured without amplitude or phase distortion. The method permits control of frequency and flow amplitude during PAlv measurement and calibration of PAlv in the absence of an active chest wall. However, it is technically complex.
This article has been cited by other articles:
![]() |
Y C G Lee, B Singh, S C Pang, N H de Klerk, D R Hillman, and A W Musk Radiographic (ILO) readings predict arterial oxygen desaturation during exercise in subjects with asbestosis Occup. Environ. Med., March 1, 2003; 60(3): 201 - 206. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Peslin and C. Duvivier Partitioning of airway and respiratory tissue mechanical impedances by body plethysmography J Appl Physiol, February 1, 1998; 84(2): 553 - 561. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |